A week ago, I presented a case on a patient with atrial pacing [AAI(R)] and a very prolonged AV delay which deteriorated to Wenckebach AV block with increasing atrial pacing rates. The low rate was 60 bpm. The same patient had a very interesting Holter 24-hour rate histogram. (Seen above).The histogram shows double counting overnight as the atrial stimulus artefacts are recognised as QRS complexes.What is more important is the lower rate of 60 bpm. There is intermittently a second lower rate of 45 bpm, which violates one of the fundamental rules of cardiac pacing; a rate lower than the programmed lower rate.Here is the ECG with atrial pacing at two lower rates

Think of the ways this can happen.

In the above figure, the patient has underlying atrial fibrillation with a slow ventricular response. Although there is ventricular pacing at 50 beats per minute (bpm), pacing will not occur until a sensing period of 1500 millisecond (ms) or 40 bpm has expired. The patient or physician could detect a pause in the pulse, confirmed on the ECG and suspect pacemaker malfunction. This can be also used with atrial pacing. In our case there was no atrial sensing.

The programmed low rate is 70 bpm. However, when the patient goes to sleep, the rate falls to 60 bpm. The algorithm may be based on an accelerometer, so that at rest, the rate falls. Other algorithms use a timer and sleep times may be automatic or need to be programmed. In all cases, activity will cancel the rest rate. In our case the pacing rate slowing is not nocturnal.These algorithms are clearly not the cause of the intermittent daytime nocturnal violation of the lower rate.Let us review the ECG tracing again.

Following the atrial stimulus artefact (or atrial sensing), there is an atrial refractory period during which there is no atrial sensing. The length of this period varies considerably between manufacturers and may be automatic depending on atrial sensing or pacing and may be, but not always, programmable.In the above case, the programmable default atrial refractory period was 350 ms. The AV delay was 340 ms and deteriorated further with increasing pacing rates. Consequently, there was a very narrow window, where the QRS could be sensed by the atrial lead. This is referred to as far-field R wave sensing and is very important with dual chamber pacing where double counting by the atrial channel can lead to mode switching.So, we have two very unusual and critical factors that together lead to atrial oversensing and “apparent” violation of the lower rate limit; a very narrow zone of open atrial sensing and far-field R wave sensing. When the R waves are detected, they are logged as atrial sensing and the pacing rate slows. This is the first time I have seen this as atrial pacing, and is not commonly used. I could find only one old reference in the literature.The issue can be very easily solved by lengthening the atrial refractory period or programming atrial sensitivity to a less sensitive (higher number) value.Remember, normally there is no R wave sensing with AAI(R) pacing. Here is an example of unipolar atrial pacing and a multifocal ventricular couplet.

Whenever I show this illustration, I am told that it is pacemaker malfunction. However, this is NORMAL atrial pacing. With bipolar pacing we usually don’t see the stimulus artefacts.
Harry Mond
October 16, 2026
Retrograde conduction is defined as conduction from an ectopic site to the atrium via the AV node or accessory pathway and referred to as VA conduction. II results in a P wave configuration which differs from an anterograde P wave, and in most ECG leads is inverted. The site from where the impulse originates can be the AV node, Bundle of His, distal conducting system or the body of the ventricles including ventricular pacing. On the ECG, the QRS complex usually precedes the inverted retrograde P wave but on occasion may follow it.
October 1, 2026
In 1970, I commenced work as a “novice” cardiologist at The Royal Melbourne Hospital. Training at that time was learning how to recognise the diastolic murmur of mitral stenosis and making sure the ECG and chest x-rays were the correct way up. As this was prior to coronary artery surgery, stenting and echocardiography, the ECG was a highly respected tool for the diagnosis of many conditions. I was fortunate to befriend a number of the gurus of ECG interpretation who taught me to look into the soul of an ECG. Today, even though the ECG is the first investigation performed by a cardiologist, the art of ECG interpretation has been lost because we depend on the new tools for cardiac diagnosis.
September 18, 2026
Dr Thura Thant asked for an opinion on this confusing ECG. By using arrows and highlight, the ECG bared its soul and gave me an answer. See if you can work it out? What do you think?